Log event window shows “Warning ENR table missing”

June 12, 2024

Noise Measurements

Symptoms / Solutions:

When running a noise project, the log event window shows the warning “Warning ENR table missing”:

FDCS will not extrapolate the ENR table, however it can interpolate. If the tuner is calibrated at frequencies beyond the ENR table range, you can cut off the frequencies outside the range of ENR table in the tuner calibration.

You can enter the ENR table by clicking the Noise Source in the setup window.

Additional Information:

The ENR table is calibrated and provided by the manufacturer.

Keywords: ENR

Last date modified:  2017 May 09