Log event window shows “Warning ENR table missing”

June 12, 2024

Noise Measurements

Symptoms / Solutions:

When running a noise project, the log event window shows the warning “Warning ENR table missing”:

FDCS will not extrapolate the ENR table, however it can interpolate. If the tuner is calibrated at frequencies beyond the ENR table range, you can cut off the frequencies outside the range of ENR table in the tuner calibration.

You can enter the ENR table by clicking the Noise Source in the setup window.

Additional Information:

The ENR table is calibrated and provided by the manufacturer.

Keywords: ENR

Last date modified:  2017 May 09

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.