This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
How do I verify that my tuner is working correctly – repeatable?
June 12, 2024
Tuner Calibration
Symptoms / Solutions:
Repeatability verification can be perform in FDCS. In the Tuners pane, right click on the tuner’s serial number –> Calibrate Tuner –> Repeatability Verification.
The Repeatability Verification window will open. Select a calibration, the VNA to use and any options, and then Run the operation.
Enter the filename to be saved and press Save. After calibration, read the S-Parameter difference value in the saved file. The S-Parameter difference should be lower than -40dB.
Keywords: Tuner, calibration, repeatability