This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
Symptoms / Solutions:
- Using FDCS to upload calibration to the tuner, the software displays the error “Calibration ID = XX contains more than 20 circles. Aborting calibration upload.”
- Recalibrate the tuner with a density no greater that Medium
- Only upload a maximum of 150 calibrations.
- Using FDCS to upload calibration to the tuner, the software stops uploading after a certain percentage.
- Connect to the tuner via telnet. Send the “format” command to delete all the existing calibrations to make room for new calibrations.
Additional Information:
The tuner has a limited on-board storage capacity and processing ability the number of calibration points and the total number of calibrations that can be stored is limited.
Keywords: Calibration, upload, error, iTuner
Last date modified: February 13, 2017
Ruggedness
Pulsed IV Systems
Device Modeling
Strictly Necessary Cookie should be enabled at all times so that we can save your preferences for cookie settings.
If you disable this cookie, we will not be able to save your preferences. This means that every time you visit this website you will need to enable or disable cookies again.
This website uses Google Analytics to collect anonymous information such as the number of visitors to the site, and the most popular pages.
Keeping this cookie enabled helps us to improve our website.
Please enable Strictly Necessary Cookies first so that we can save your preferences!