Error during Wave Load Pull setup using the PNA-X

June 12, 2024

Measurements

Symptoms / Solutions:

Wave Load Pull setup back to back verification on a through with the Load and Source Tuners initialized gives power gain close to 0 dB as expected. However, verification with Gamma load values other than 50Ω gives wrong power gain values.

To obtain proper receiver measurements using the PNA-X, ensure that the file IFMUXCAL.txt, located in C:\Program Files\Agilent\Network Analyzer is correct (R1 to R4 must be similar). This file contains receiver offsets used for IF calibration. If corrupted, rename it and put all the jumpers in place. With nothing connected to the PNA-X ports, run the application ifcal.exe located in C:\Program Files\Agilent\Network Analyzer\Service\ directory. The PNA-X will generate a new IFMUXCAL.txt file with correct numbers.

 

An example of a correct IFMUXCAL.txt file appears as follows:

IFMUX CAL
R1
5.9831350906305
0.815871697708853
R2
5.94147855637884
-1.1512360736319
R3
5.92475533872457
-0.525978706721356
R4
5.90785275611497
-2.33257454708033
R1
5.97201435283602
0.598166088938812
R2
5.96967768627871
0.364352395744098
R3
5.9022240418708
0.386678890694407
R4
5.97775613234385
0.00946239937570624

Figure 1 Valid IFMUXCAL file

 

Additional Information: Additional information on the Wave Load Pull setup can be found in the Wave-based Load Pull Manual in \Focus Microwaves\Load Pull Explorer\Manuals\Applications

Keywords: Wave Load Pull, Back to Back verification, Power gain, IFMUX cal

Last date modified: October 12, 2017